Fig. 12
3D reconstruction PC
NOTE
Disconnect the system from the power supply by pulling the mains plug out of the mains
socket.
The syngo software and the backup/restore disk are kept in the monitor trolley logbook.
Unplug all connectors on the PC.
Unscrew the dongle from the printer port of the PC and reinstall it on the new PC
(see Fig. 12).
Replace the PC.
Reconnect all connectors.
Perform the startup of the 3D reconstruction PC as described in the Startup Instructions,
Chapter 4.
Then insert the backup / restore disk (in the logbook compartment of the
SIREMOBIL Iso-C) into the disk drive of the PC.
Start the syngo service software.
Select the Backup / Restore menu and restore the data.
Perform a function check of the PC.
Perform the IQ quick test.
Check the DICOM functionality of the PC, if applicable.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Power out (switched
Power in
Mains power switch
Mouse port
Keyboard port
Printer port (dongle)
Serial port
Ethernet LAN port
Audio out Audio in / Micro
Video graphic port
CAN port
SCSI adapter (MEMOSKOP)
Relais board
SCSI adapter
Only applies to SIREMOBIL Iso-C 3D.
SPR2-230.061.01