6 - 38
Fig. 12
3D reconstruction PC
NOTE
•
Disconnect the system from the power supply by pulling the mains plug out of the mains
socket.
•
The syngo software and the backup/restore disk are kept in the monitor trolley logbook.
•
Unplug all connectors on the PC.
•
Unscrew the dongle from the printer port of the PC and reinstall it on the new PC
(see Fig. 12).
•
Replace the PC.
•
Reconnect all connectors.
•
Perform the startup of the 3D reconstruction PC as described in the Startup Instructions,
Chapter 4.
•
Then insert the backup / restore disk (in the logbook compartment of the
SIREMOBIL Iso-C) into the disk drive of the PC.
•
Start the syngo service software.
•
Select the Backup / Restore menu and restore the data.
•
Perform a function check of the PC.
•
Perform the IQ quick test.
•
Check the DICOM functionality of the PC, if applicable.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Power out (switched
Power in
Mains power switch
Mouse port
Keyboard port
Printer port (dongle)
Serial port
Ethernet LAN port
Audio out Audio in / Micro
Video graphic port
CAN port
SCSI adapter (MEMOSKOP)
Relais board
SCSI adapter
Only applies to SIREMOBIL Iso-C 3D.
SPR2-230.061.01
Rev. 13
Page 38 of 48
12.04
CS PS 24
6
Siemens AG
Medical Solutions