Saving Current Trace Files Directly After Display Of An Error - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Loading the operating software

Saving current trace files directly after display of an error

Each time the system is switched on a trace file SPETrace_00.TRC is created and writ-
ten and filed under the drive C:\Trace\SPETrace_00.TRC.
The current error trace can be saved via "local service" Utilities, source, "escape to OS".
Insert an empty formatted floppy disk in drive A of the Syngo PC.
In Syngo select the "local service" menu and enter the password.
Activate the "Utilities" button.
Under Source activate "Escape to OS".
In the Parameters line enter "copy C:\traces\*.trc A:" and then activate the "Go" button.
The trace files are copied to floppy disk.
Wait until the confirmation is displayed in the window.
Fig. 2
Siemens AG
Medical Solutions
SPR2-230.061.01
Rev. 13
12.04
Page 3 of 6