Loading the operating software
Saving current trace files directly after display of an error
•
Each time the system is switched on a trace file SPETrace_00.TRC is created and writ-
ten and filed under the drive C:\Trace\SPETrace_00.TRC.
The current error trace can be saved via "local service" Utilities, source, "escape to OS".
•
Insert an empty formatted floppy disk in drive A of the Syngo PC.
•
In Syngo select the "local service" menu and enter the password.
•
Activate the "Utilities" button.
•
Under Source activate "Escape to OS".
•
In the Parameters line enter "copy C:\traces\*.trc A:" and then activate the "Go" button.
The trace files are copied to floppy disk.
•
Wait until the confirmation is displayed in the window.
Fig. 2
Siemens AG
Medical Solutions
SPR2-230.061.01
Rev. 13
12.04
Page 3 of 6
SIREMOBIL Iso-C
CS PS 24
4 - 3
4
System Manual