Prerequisites
Required documents
•
SIREMOBIL Iso-C circuit diagram
•
Circuit diagram for monitor X2080
•
SIREMOBIL Iso-C logbook
•
For the laser targeting device where used:
For SIREMOBIL Iso-C 3D additional
•
SIREMOBIL Iso-C 3D
•
Image Quality Quick Test
Required tools, measurement and auxiliary devices
•
DMM
•
Oscilloscope > 50 MHz e.g. TFluke Combi Scope PM 3390A
•
Dose measurement device
•
Safety tester
•
Service PC with WINDOWS 3.1., 3.11 or 95 operating system
•
Service software (refer to SIREMOBIL Iso-C logbook)
•
Connection cable, Service-PC - SIREMOBIL Iso-C
•
1 set of resolution tests
•
1 set of radiation filters
•
Center cross
•
Heat conducting paste WPS
•
Special purpose oil (Optimol Optipit)
•
Special purpose oil (Slic Pac PTFE)
•
Sealing compound
•
Standard service equipment*
•
1 set Allen keys*
•
200 N spring scale
•
Torque wrench 20 Nm to 100
Nm
•
Torque wrench 3 Nm to 20 Nm
•
A 30 mm socket key (is required for adjusting the brake)*
For SIREMOBIL Iso-C 3D additional
•
Calibration Phantom Iso-C 3D (without navigation interface)
or Calibration Phantom Iso-C 3D (with navigation interface)
•
Luminance meter
Siemens AG
Medical Solutions
Adjustment Instructions / Laser targeting
device RXR2-130.032.01...
Startup Instructions SPR2-230.815.01...
SPR2-230.037.01 Revision 05 or higher
e.g. Fluke 8060A
e.g. PTW-DIADOS
e.g. BENDER Unimet 1100 ST
e.g. SMfit light field lum. meter
SPR2-230.061.01
Rev. 13
12.04
1
97 02 101
99 00 861
97 17 612
51 38 727
e.g.
99 00 440
e.g.
28 71 820
e.g.
97 98 596
e.g.
96 60 051
e.g.
20 48 650
55 07 525
55 07 517
20 49 716
e.g.
44 15 113
e.g.
80 86 159
e.g.
34 24 553
71 39 947
75 51 620
88 81 281
Page 1 of 18
SIREMOBIL Iso-C
CS PS 24
1 - 1
1
1
System Manual