6 - 32
Language
Display of realtime clock
Display of rotated images
Monitor zoom
Monitor split
Monitor text
Maximum image number
Disable doserate selection
in user setup
Unit serial number
Product name
Software backward compati-
ble
Default program change-
able
Create external disk
Enable Bridge
Delay 3D-PC/ DICOM
Bridge
Interface to 3D-PC
Default LUT
Default Edge
Default LUT for Subtraction
Subtraction K-Factor,
phase A
Subtraction K-Factor,
phase B
Subtraction Time of phase A Program "4,5".
Roadmap K-Factor, phase A Program "32" .
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Set for language of the country.
Configure Yes. Select No only upon customer request.
Configure Yes.
Configure No.
Configure Yes.
Configure Yes.
Configure 900.
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are vis-
ible. Otherwise set to "Yes".
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Program "SIREMOBIL Iso-C".
Configure No.
Default: Yes.
Depending on country specific regulations configure Yes or
No.
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Only available of if option "3D Reconstruction" is installed.
Configure 120.
Only available if option "3D Reconstruction" is installed.
Configure yes.
Configure 2. Only upon customers request program a
different LUT curve.
Configure 1. Only upon customers request program a
different Edge enhancement curve.
Configure 1. Only upon customers request program a
different LUT curve.
Program "32". Program a different value only upon cus-
tomer request.
Program "MD 2". Program a different value only upon cus-
tomer request.
Program a different value only upon customer request.
Program a different value only upon customer request.
SPR2-230.061.01
Rev. 13
Page 32 of 48
12.04
CS PS 24
Siemens AG
Medical Solutions