42
3.2.5

Handles

SIM
Attachment
Verify that the support arm and collimator handles are undamaged and securely at-
tached. Tighten screws and replace defective parts as necessary.
Fig. 8:
3.2.6

Collimator adjustment knobs

SIM
Attachment
Verify that the collimator adjustment knobs are properly seated.
3.2.7

Arm system and single tank

3.2.7.1
Moving the arm system
SIM
Moving the arm system
Move the support arm into all the usual positions. The arm should not move up or down
after you release it. (To adjust the friction brakes, refer to the "Replacement of parts" in-
structions).
SIM
Moving the single tank holder
The single tank holder must be easy to turn and maintain any position.
SIM
Moving the single tank
The single tank must be easy to turn and maintain any position.
MOBILETT XP Digital
Collimator handle
SPR8-230.831.30.03.02
11.05
Inspection and maintenance