NOTE
NOTE
MOBILETT XP Digital
General maintenance information
File the patient leakage current report in the "Certificates" register
no. 9 in the system binder.
A new report must be created if the patient leakage current mea-
surements are not documented.
A new patient leakage current report can be found in the startup in-
structions SPR8-230.814.30.03. Separate this report from these in-
structions, fill it out, and file it in the "Certificates" register no. 9 in
the system binder. If values are newly determined, they must be re-
corded as first measured values. The evaluation is omitted in this
case.
SPR8-230.831.30.03.02
11.05
Page 30 of 74