32
Test report/
Technical safety checks
2.3 Are undamaged accessories being
used?
2.4 System radiation protection devices
present and undamaged? (No on-site
radiation shield)
2.5 Are the required operator documents
complete, present and legible?
2.6 Are all warning labels in place and
recognizable?
2.7 Are the operating symbols, light indi-
cators and button labels OK?
3. Electrical checks
3.1 Measurement of ground wire resis-
tance
3.2 Leakage current or equivalent leak-
age current measurement
3.2.1 Leakage current measurement
3.2.2 Equivalent leakage current mea-
surement
3.3 Patient leakage current or equivalent
patient leakage current measurement
MOBILETT XP Digital
General maintenance information
SPR8-230.831.30.03.02
11.05
Maintenance instructions for the
check/comments
3.1.1 Damage, check:
SIM DAP (optional)
SIM Remote control (optional)
3.3.3 Check of the radiation indicator,
checks:
SIE Radiation indicator
SIE Ready indicator
SIE Acoustic signal
3.3.4 Manual termination of exposure,
checks:
SIE "ERR 39" display
SIE Acoustic signals
3.1 Visual inspection, check:
SI Customer documentation
Ask the customer for additional required
operator documents and check them for
completeness, availability, and legibility.
3.1 Visual inspection, check:
SIM Signs
3.3.2 Control panel displays, checks:
SIE kV/mAs 7-segment displays
SIE kV/mAs displays according to operating
instructions
n.a.
3.10 Protective conductor test, check:
SIE Protective conductor test
n.a.
3.11 Leakage current measurement, check:
SIE Leakage current measurement
n.a.
n.a.
Page 32 of 74
CS SD 24
Siemens AG
Medical Solutions