Collimator; Lamp Replacement - Siemens MOBILETT XP Digital Maintenance Instruction

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

52
3.4

Collimator

3.4.1

Lamp replacement

PMP
Annual replacement of the collimator lamp
The collimator lamp should be replaced every 12 months.
U
Tools and auxiliary materials required
- Rope for securing the support arm (approx. 4m long)
- Allen wrench set
- Screw driver set 1-6 (Phillips head and slotted head screws)
24V/100W lamp - Siemens order no.: 08392016
NOTE
Required work steps:
Switch off the system and possibly disconnect the power plug.
Use a rope or tightening belt to secure the system support arm against lifting
(B/Fig. 18 / p.
Remove the two adjustment knobs of the collimator (2 mm Allen wrench)
(C/Fig. 19 / p.
Disconnect the plug of the DAP measurement chamber (if present).
Open the collimator cover
Remove the protective cover from the lamp
Replace the lamp
To reassemble, repeat the above steps in reverse order.
PMF
Light localizer function
Turn the system ON and test the function of the light for all 3 light localizer buttons.
MOBILETT XP Digital
Do not touch the lamp with bare hands. Touch the bulb only with a
clean, dry cloth.
53).
The single tank should rest securely on a table or the like.
53).
Note the position of the knobs during removal.
(1/Fig. 20 / p.
Turn the cover and remove the connections: DAP plug; light switch, ground wire
(Fig. 21 / p.
54).
(6/Fig. 23 / p.
The light filament of the lamp must be positioned directly over the rectangular
light aperture.
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
54).
(4/Fig. 22 / p.
55).
Page 52 of 74
CS SD 24
54).