Collimator
3.4.1
Lamp replacement
PMP
Annual replacement of the collimator lamp
The collimator lamp should be replaced every 12 months.
U
Tools and auxiliary materials required
- Rope for securing the support arm (approx. 4m long)
- Allen wrench set
- Screw driver set 1-6 (Phillips head and slotted head screws)
24V/100W lamp - Siemens order no.: 08392016
NOTE
Required work steps:
•
Switch off the system and possibly disconnect the power plug.
•
Use a rope or tightening belt to secure the system support arm against lifting
(B/Fig. 18 / p.
•
Remove the two adjustment knobs of the collimator (2 mm Allen wrench)
(C/Fig. 19 / p.
•
Disconnect the plug of the DAP measurement chamber (if present).
•
Open the collimator cover
•
Remove the protective cover from the lamp
•
Replace the lamp
•
To reassemble, repeat the above steps in reverse order.
PMF
Light localizer function
•
Turn the system ON and test the function of the light for all 3 light localizer buttons.
MOBILETT XP Digital
Do not touch the lamp with bare hands. Touch the bulb only with a
clean, dry cloth.
53).
The single tank should rest securely on a table or the like.
53).
Note the position of the knobs during removal.
(1/Fig. 20 / p.
Turn the cover and remove the connections: DAP plug; light switch, ground wire
(Fig. 21 / p.
54).
(6/Fig. 23 / p.
The light filament of the lamp must be positioned directly over the rectangular
light aperture.
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
54).
(4/Fig. 22 / p.
55).