General maintenance information
Test report/
Technical safety checks
6. Product-specific checks
7. Check results/evaluation:
Siemens AG
Medical Solutions
SPR8-230.831.30.03.02
11.05
Maintenance instructions for the
check/comments
3.2.5 Handles, check:
SIM Attachment
3.2.6 Collimator adjustment knobs, check:
SIM Attachment
3.2.7.2 Arm system attachment, check:
SIM Screws and lock nuts
3.2.7.3 Arm connector, check:
SIM Screws and nuts
3.2.7.4 Adjusting screw and cantilever,
checks:
SIM Lock nuts
SIM Attachment of the base
3.2.7.8 Single tank holder, checks:
SIM Holder screw connections
SIM Single tank screw connections
3.6.1 DAP measuring system (optional),
check:
SIE Function of the DAP measuring system
3.6.2 Remote control (optional), check:
SIE Function of the remote control
Evaluate the long-term trends of the ground
wire resistance and the leakage current by
comparing the current measuring values to
those of the preceding technical safety
checks. A sudden or unexpected increase in
the measured values may indicate a
safety-related defect - even if the limit val-
ues are not exceeded.
Enter the results of the check.
Have the operator, or a person authorized
by the operator, sign the TSC protocol.
Hand the TSC protocol over to the operator,
or a person authorized by the operator.