Displaying The Control Panel; Checking The Radiation Indicator - Siemens MOBILETT XP Digital Maintenance Instruction

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

50
3.3.2

Displaying the control panel

PMF
Mains display
The yellow line voltage control LED must be lit when the line voltage plug is plugged in
(3/Fig. 17 / p.
System ON.
PMF
kV/mAs default values
SIE
kV/mAs segment displays
The kV/mAs default values must display and be completely legible (take the default val-
ues from the start-up protocol).
PMF
Ready indicator
The green ready LED
ON".
SIE
kV/mAs displays according to operating instructions
Select all kV/mAs values.
3.3.3

Checking the radiation indicator

Fig. 17: Control and display
SIE
Radiation indicator
SIE
Ready indicator
SIE
Acoustic signal
System is switched on.
Set 81 kV and 10 mAs.
X
Initiate an exposure.
The radiation indicator
time an acoustic signal sounds and the "ready" light goes out.
MOBILETT XP Digital
50).
(2/Fig. 17 / p. 50)
(1/Fig. 17 / p. 50)
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
should light up within 15 seconds of "system
lights up during the exposure; at the same
Page 50 of 74