50
3.3.2
Displaying the control panel
PMF
Mains display
•
The yellow line voltage control LED must be lit when the line voltage plug is plugged in
(3/Fig. 17 / p.
•
System ON.
PMF
kV/mAs default values
SIE
kV/mAs segment displays
•
The kV/mAs default values must display and be completely legible (take the default val-
ues from the start-up protocol).
PMF
Ready indicator
•
The green ready LED
ON".
SIE
kV/mAs displays according to operating instructions
•
Select all kV/mAs values.
3.3.3
Checking the radiation indicator
Fig. 17: Control and display
SIE
Radiation indicator
SIE
Ready indicator
SIE
Acoustic signal
•
System is switched on.
•
Set 81 kV and 10 mAs.
•
X
Initiate an exposure.
The radiation indicator
time an acoustic signal sounds and the "ready" light goes out.
MOBILETT XP Digital
50).
(2/Fig. 17 / p. 50)
(1/Fig. 17 / p. 50)
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
should light up within 15 seconds of "system
lights up during the exposure; at the same
Page 50 of 74
CS SD 24
Siemens AG
Medical Solutions