Technical Safety Checks (Tsc) - Siemens MOBILETT XP Digital Maintenance Instruction

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

General maintenance information
2.10

Technical Safety Checks (TSC)

Abbreviation: TSC = Technical Safety Checks
NOTE
Test report/
Technical safety checks
1. Identification data
2. General checks
2.1 Visual inspection: Is there any dam-
age affecting safety?
2.2 Are all cables and cable guides
secure and without any visible damage?
Siemens AG
Medical Solutions
Within the purview of DIN VDE 0751-1, the operator of medical en-
gineering products has to perform technical safety checks at reg-
ular intervals.
The checks listed in these maintenance instructions include all
technical safety checks required by DIN VDE 0751-1.
A separate protocol with the print number SP00.000.834.01... is
available for each technical safety check.
Within the purview of DIN VDE 0751-1, we recommend document-
ing the maintenance results both in the maintenance protocol and
in the TSC protocol. The protocols should be filled out completely
and handed over to the client after maintenance is complete.
The table below assigns each technical safety check listed in doc-
ument SP00.000.834.01... to the corresponding maintenance work.
SPR8-230.831.30.03.02
11.05
Maintenance instructions for the
check/comments
n.a.
n.a.
3.1.1 Damage, checks:
SIM Covers
SIM Detector/cassette holder
SIM Control and display panel
SIM Release cable
SIM Single tank
SIM Collimator
3.2.9 Power cable, checks:
SIM Power cable check
SIM General cable winch information
3.2.9.1 Power plug, check:
SIM Power plug check
3.2.7.5 Cable harness for the arm system,
check:
SIM Damage-free
Page 31 of 74
CS SD 24
31
MOBILETT XP Digital

Hide quick links:

Advertisement

Table of Contents
loading