Inspection and maintenance
3.10
Protective conductor test
SIE
Protective conductor test
Observe the protective conductor resistance test information in these instructions.
•
Perform the protective conductor test with all covers closed, according to
ARTD-002.731.17....
The protective conductor resistance must not exceed 0.2 Ohms.
•
The values determined, including the measuring points, must be recorded and as-
sessed in the protective conductor resistance report.
•
The measuring procedure and the measuring device used (designation and serial num-
ber) must also be documented.
NOTE
NOTE
Siemens AG
Medical Solutions
File the protective conductor resistance report in the "Certificates"
register no. 9 in the system binder.
A new report must be created if the protective conductor resis-
tance measurements are not documented.
A new protective conductor resistance report can be found in the
startup instructions SPR8-230.814.30.03. Separate this report from
these instructions, fill it out, and file it in the "Certificates" register
no. 9 in the system binder. If values are newly determined, they
must be recorded as first measured values. The evaluation is omit-
ted in this case.