8041A

Message
Display
Signal
Indicator
indicator
Panel
lamps
flashing
*. No longer available
Chapter 3 - Tests and Error Messages
Cause
Invalid telemetry mode.
Incorrect transducer
connected into
transmitter. (Only
Ultrasound and Toco
transducers can be used).
Transducers are
connected to the front
panel of the monitor.
Solution
Check the cable from the telemetry receiver and,
if necessary, replace it.
Check that the transducer is compatible with
Series 50T Fetal Telemetry System.
Disconnect the transducers from the monitor or
switch off the telemetry receiver.
Error Messages