3.2.6.3 Action to be taken when software crashes
In case of a software crash, the system can be recovered by reinstalling the ultrasound
scanner software. The time required is approximately 30 Minutes
Procedure to reinstall software..
Switch on the unit and insert System CD into the ultrasound scanner's
(1) During the boot phase. A white bar will appear on the screen indicating that the unit is
booting from the System CD when displaying HITACHI Logo mark.
(2) When the start screen appears, ( Approximately 5 Mins ) depress "Ctrl + Alt + E" key
This operation needs to be executed just when the start screen is displayed. If not the e
ultrasound scanner completes the boot up and displays the US image.. In such a case,
turn OFF the system power and retry from the beginning.
NOTICE: The activation time of Explorer is long. Don't push "Ctrl + Alt + E" key
NOTICE: Except for case of invalidating STBY SW, don't turn off by breaker SW..
frequently.