3.2.8
Cautions in replacing EPI PCB
(1)
When replacing the EPI  PCB, be sure to turn off the breaker switch located at left side of
the system.
(2)
EUB-5500 is numbered as individual ID number for each system, and this ID number is
recorded in ID chip which is equipped on the EPI PCB. (Location on the PCB C8 : Refer
to the figure below.)
In case no ID chip is equipped or an incorrect ID chip for different system is equipped,
the system does not operate correctly. When replacing EPI PCB, make sure to
re-equip this ID chip onto the new PCB to be replaced.
CNF5
CNF8
C8
EPI P.C.B.
Parts side
3 - 39
L1E-EA0229