16. Is
the YO Main Coil Drive Correct?
Check the YO MAIN COIL DRIVE. This
can be
done
by testing the
SENSE
voltage, TP4
on the ASS
YO driver assembly. (The SENSE
voltage
is
related to the
YO frequency
by
-2.4
V/GHz.)
If
the
YO
main coil
is
greatly mistuned, no
YO
signal will be present.
If
the
SENSE
voltage (hence the YO
MAIN COIL DRIVE current)
is
correct
to within
±1V,
troubleshoot the YO;
if
not,
proceed
to
question
17.
17. Is
the PRETUNE Voltage Correct?
This portion of the troubleshooting
guide
is
a
branch from question 3, question
10,
and question
16.
Check
the
PRETUNE
voltage.
PRETUNE
has
a
sensitivity
of
-2.5
V/GHz
of YO frequency.
If
the
PRETUNE
voltage
is
correct, troubleshoot the
ASS
YO driver (OFFSET
and
GAIN)
and the
A47
sense resistor assembly.
If
PRETUNE
is
not correct, proceed to question
18.
18. Is
the VSWP Voltage Correct?
Check
that the
VSWP
voltage
is
zero
volts.
If it is
correct, suspect the AS4
YO pretune/delay
compensation assembly.
If
VSWP
is
much different than
zero
volts,
proceed to question
19.
19.
Is LVSX Low?
If
the
VSWP
voltage
is
much different than zero
volts,
it will
affect the
PRETUNE
voltage only
if
the
AS4 YO
pretune/delay compensation assembly
is
also not operating correctly. If LVSX (low
volt¬
age sweep
disable)
is
high or intermittent, troubleshoot the
ASS digital interface
assembly
(in the
controller
section).
If
LVSX
is
low (that should be the case for
CW/MANUAL), the trouble
is in
the
AS4
YO pretune/delay compensation assembly.
In
either case, also troubleshoot
VSWP
on the
ASS
sweep generator assembly.
YO UNLK
-
SWEPT MODE
To troubleshoot
in
the
swept mode with the YO unlocked, perform the
following steps:
1.
Is
CW/MANUAL Mode Correct?
If the
YO UNLK error condition can
be
made
to
occur
in
the
CW/MANUAL
mode, go to the YO
UNLK
-
CW/MANUAL
troubleshooting section.
It will
also
be helpful
if
the
YO UNLK
condition
can be made to occur
in
the SINGLE SWEEP mode,
for
this will hold
the YO
loop (error) condition
constant while troubleshooting.
If
the error
condition can only be
made
to
occur
In
the
SWEPT
mode and
CONTINUOUS SWEEP
mode, the problem
will
be more
difficult to
solve. Proceed to question
2
in
this section.
There are
several malfunctions that could cause
a
YO loop unlock
in
only the swept mode
these relate
to
the function and operation of
VSWP,
LKICK, HLEY
(high lock
enable YIG
oscillator), and possibly
VCOMP,
LVSX, and LYSP. Most of the potential causes of the error are related
to timing,
mainly at
the start
of sweep.
D-18
YO
Loop Assembly-Level
Troubleshooting
HP 8340B/41B