INCORRECT SWEEP
If
the instrument sweeps, as defined above, but the YO
does
not
sweep
or the frequency limits
are
incorrect,
check
the
two sweep ramps from
the ASS sweep
generator assembly, 20-30 SWP (TP8) and
VSWP
(TP10).
The
sensitivities
of these
lines are
discussed below.
YO sweeps
widths
>5
MHz
are derived from VSWP. Note that the instrument sweep width, delta
F, is
related to the
YO sweep
width by the harmonic
number
of the band being used. For example,
if
the
instrument
is
sweeping
from
8
to
10
GHz, the
instrument
sweep width
is
2
GHz.
However,
this
is
the
second harmonic of the YO
the
YO
is
actually
sweeping from
4
to
5
GHz;
a
sweep
width of
1
GHz.
The sensitivity of VSWP
is
+2
volts/GHz
of
YO sweep
width.
In
the
example
above,
VSWP
should be
a
ramp
starting at
zero and
going to
+2
volts.
If
VSWP
is
correct, look
at
PRETUNE (TP3)
on the
A54 pretune/delay
compensation assembly.
It
has
a
sensitivity of
—2.5 volts/GHz of YO frequency.
In
the foregoing
example where
the
YO swept from
4
to
5
GHz, PRETUNE
should go from
-10
to
-12.5
volts.
If
PRETUNE
is
correct, look at
SENSE (TP4)
on
the A55 YO driver assembly. This point has
a
sen¬
sitivity of
about —2.34 volts/GHz of YO frequency.
If
this
is
correct, the problem
is
most likely
in
the
A44 YIG
oscillator assembly.
YO sweep
widths
^5
MHz
are derived from 20-30 SWP.
The
sensitivity of
20-30 SWP
is
related to
YO
sweep width but changes depending upon
the particular
sweep
width selected, as shown
in
Table
D-2.
Table D-2. Sensitivity of 20-30
SWP
Line
Sweep Width
5
MHz
^
YO sweep
width
>
500
KHz
500 KHz
2: YO
sweep
width
>
100 KHz
100 KHz
s
YO sweep width
>
10
KHz
10 KHz
^
YO sweep
width
>
1
KHz
5
KHz
^
YO sweep
width
>
500 Hz
500 Hz
2:
YO sweep width
>
100 Hz
Sensitivity
500 KHz/volt
50 KHz/volt
10
KHz/volt
1
KHz/volt
500 Hz/volt
50 Hz/volt
Verify that
VSWP
is
not
sweeping more than 0.0
to
15.0
mV
and that
20-30 SWP
is
correct.
If
both of
these
are
correct, the problem
is
most likely
in
the
20-30
loop.
HP 8340B/41B
Sweep
Generator
Assembly-Level Troubleshooting