SHIFT M4
(HP-IB:
SHM4)
SHIFT M4
activates 32 service diagnostic routines that test some of the DACs and much of the control
circuits on the
A27
level control,
A28 SYTM
driver,
A57 marker bandcross,
and ASS
sweep generator
assemblies. This diagnostic also allows the
results of the self test, run at
power on and after
instru¬
ment preset, to be displayed
in
the front panel ENTRY DISPLAY.
The DAC
tests are not
exhaustive
(a small error does not
generate
a
failure indication), but
if
a
test
fails,
the indication should direct you
to
a
specific device or circuit path.
While
the tests are running, the
DIAGNOSTIC TEST
IN
PROGRESS
appears
in
the front panel
ENTRY
DISPLAY. When the
tests finish,
TEST:? FULL DIAGNOSTIC
and then
PASS
or FAIL appear. Pass
indicates
that
all tests related to this diagnostic
passed.
Fail indicates
that one
or
more
tests failed.
See
the controller section functional group for details on the
SHIFT
M4
diagnostics.
SHIFT M5
(HP-IB:
SHM5)
SHIFT M5
turns
off
any of the above diagnostic routines, and restores the
displays to their
previous
condition.
SHIFT
RF (HP-IB:
SHRF)
SHIFT RF
disables the ALC (automatic leveling control) to allow direct control of the linear modulator
circuit.
When
SHIFT
RF
is
engaged, there
is
no limit on the minimum pulse repetition
frequency
1.
Press [SHIFT] [RF]
POWER SEARCH:X.XXdB appears
in
the entry display (X.XX
is
the last-entered value).
2.
Place the instrument
in
CW, or
in
pulse modulation
mode
with pulses
wider than
2
(AS.
3.
Using the step keys, the data entry keyboard, or the front panel knob, set the desired power level.
The POWER
dBm display shows
the actual
power when
the
instrument
is in
the
CW
or pulse
modulation mode.
SHIFT
RF
can also be
used as
a
diagnostic function of the ALC
circuits.
SHIFT METER
(HP-IB:
SHA3)
This function
bypasses the ALC
to allow direct control of the linear modulator
circuit. In this
mode,
there
is
no
limit
on the minimum pulse repetition
frequency.
1.
Press
[SHIFT] [METER].
ATN: X-xx
dB,
MOD; x.x dB appears
in
the
ENTRY DISPLAY
(x
is
the last-entered value).
2.
Place the instrument
in
CW, or
in
pulse modulation mode with pulses wider than
2
ps.
HP 8340B/41B
Assembly-Level Service Front Panel Diagnostics
A-29