Calibration Constants Description - HP 8340B Manual

Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

CALIBRATION CONSTANTS DESCRIPTION
Table A-3
lists
the calibration constants with
a
description of each.
The values
in
the Range column
show
the amount of adjustment possible with the
cal
constant.
The
numbers shown are the only
numbers you
should enter. You can
enter values
out of the given range, but they do not
represent
the
values used by the instrument.
The
column labeled Significance gives the
units
associated
with the cal constant. For
example, the
value entered for
LOCK DWELL TIME represents
the number of milliseconds the instrument waits,
after phase
lock, at
the beginning of
a
sweep, or at
bandcross, before
it
continues
its
sweep. Another
example
is
the
ALC
9
GHz SLOPE, for
each count,
a
power compensation of 0.0025
dB/GHz
is
added
to
the RF output [if
10
is
entered,
a
0.025 dB/GHz
slope
(10
times 0.0025
dB/GHz)
is
added for
a
sweep
within
the
9
to 20 GHz
range].
Five
of the Cal Constants
[PRESET OPTION
(#56), HP-IB
ADD
&
LOCK (#57), RETRACE DWELL
(#58),
ATTENUATOR CONFIG
(#59), and
CONFIGURATION
(#60)]
use their
bit
configuration to store
more than one
piece of information. For
example,
the HP-IB
ADDRESS
sets up
the
HP-IB address and
allows or disallows the
SHIFT
LOCAL, SHIFT SAVE, and SHIFT RECALL
functions.
If
a
you
wish
to
set
up an
HP-IB address
of 19, disallow the
SHIFT LOCAL
and
RECALL
functions, and allow the
SHIFT SAVE
function, enter
179 for Cal Constant #57. You
find
the appropriate calibration constant
(179
in
this
example)
by adding the appropriate
calibration
constants for the
individual functions:
19
(HP-IB address)
32
(disallows
SHIFT LOCAL)
0
(allows
SHIFT SAVE)
128 (disallows
SHIFT RECALL)
"179
The other four
cal
constants work the same way.
A-16
Assembly-Level Service Calibration Constants
HP 8340B/41B

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents