2.
Press [INSTR
PRESET] and
check for the following sweep
indications:
a.
Check
front panel
SWEEP
LED. The LED should be
blinking.
The sweep LED,
which
is
turned on at start of sweep,
off
at each bandcross, and at
end-of-
sweep,
is
controlled by
LSPLD (low sweep
LED). The
sweep LED
is
turned on when HSP (high
start sweep)
is
low,
LRSP (low reset sweep)
is
high,
and LBX (low bandcross)
is
high.
b.
Check the
SWEEP OUT
signal.
The sweep outputs come from the A57 marker/bandcross assembly, and are generated from
the
MKR RMP.
For
a
detailed description of the
SWEEP OUT
or
MKR RAMP waveform,
refer
to
the sweep
generator/YO
loop functional group.
If the
sweep out waveform
is
normal, the
ASS
assembly ramp generator
is
good, the
WSPTM
(write
sweep
time) strobe
is
generated, and LRESET
(low reset)
is
working.
c.
Check rear
panel O.SV/QHz
waveform.
The 0.5V/GHz
signal
comes from the A28 SYTM
driver
assembly, and
is
generated by the
PRETUNE
signal, which
is
generated from VSWP.
If
the rear panel O.SV/GHz signal
is
normal,
VSWP
and
PRETUNE
are working.
NOTE:
For sweep widths
<5
MHz (20/30
sweeps), 0.5V/GHZ
is
fixed.
SPECTRAL PURITY
/
HARMONICS
(17,18)
NOTE:
In this
section, spectral
purity
is
considered
a
phase
noise problem and spurious
responses
are considered to
be discrete harmonics.
Phase Noise
This procedure
assumes the
instrument
Failed
the Single Sideband Phase Noise
test,
in
Section
IV.
1.
If
the instrument failed
the test at offset frequencies
less than
300
Hz, the problem
is
most likely
the 100
MHz
Reference section. Refer
to the
reference
loop
-
M/N
loop.
a.
Replace
the
400 MHz
input
to A31
M/N
phase detector
(J1)
using
a
very stable source. Use the
HP 8662A frequency standard
as the HP
8340B/41B
frequency standard EXT input.
b.
Repeat
the failed test.
If
the HP
8340B/41B now passes the
test, the problem
is in
the
HP
8340B/41B
reference
loop.
2.
If
the instrument failed the phase noise test at offset Frequencies from
300
Hz to 50 kHz,
the
problem
is
most
likely the
M/N
loop.
a.
You can
use an external
source
to
replace the
M/N
input to the YO loop, but the external
source phase
noise would have to be much better than the
HP 8340B/41B phase noise
to
eliminate
the
YO
Loop.
b.
Replace
the reference loop
400
MHz
input to the
M/N
loop
as described
in
step
1.
HP 8340B/41B
Assembly-Level Service Overall Instrument Troubleshooting