4
Fig. 1
Installation
Removing the covers
•
Disconnect the SIREMOBIL line voltage plug.
•
Remove the three cover plates from the basic unit.
•
Remove the SIREPHOS cover.
First, carefully remove the adhesive strip on the side panels of the cover.
Installing the area dose measurement chamber
•
Pull the measurement chamber connector (1/Fig. 1) out of the cable tie.
•
Place the area dose measurement chamber on the SIREPHOS as shown in Fig. 1.
The "Focus" label must point toward the X-ray tube.
•
Secure the area dose measurement chamber to the SIREPHOS with the enclosed
screws.
•
Plug the measurement chamber connector (1/Fig. 1) into the measurement chamber.
Installing the measurement board D16
•
Install measurement board D16 according to Fig. 2.
•
Plug the two connectors (2/Fig. 2) into PC board D16.
•
Connect the enclosed ribbon cable to D1.X9 and D16 (3/Fig. 2).
SIREMOBIL Compact
Installation and Setting Instructions
1
Register 3
RXR2-130.033.01
Rev. 04
2
3
2
Fig. 2
Page 4 of 10
04.02
CS PS 24
1
1
1
1
Siemens AG
Medical Engineering