Installation - Siemens SIREMOBIL Compact Installation And Setting Instructions Manual

Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

4
Fig. 1

Installation

Removing the covers
Disconnect the SIREMOBIL line voltage plug.
Remove the three cover plates from the basic unit.
Remove the SIREPHOS cover.
First, carefully remove the adhesive strip on the side panels of the cover.
Installing the area dose measurement chamber
Pull the measurement chamber connector (1/Fig. 1) out of the cable tie.
Place the area dose measurement chamber on the SIREPHOS as shown in Fig. 1.
The "Focus" label must point toward the X-ray tube.
Secure the area dose measurement chamber to the SIREPHOS with the enclosed
screws.
Plug the measurement chamber connector (1/Fig. 1) into the measurement chamber.
Installing the measurement board D16
Install measurement board D16 according to Fig. 2.
Plug the two connectors (2/Fig. 2) into PC board D16.
Connect the enclosed ribbon cable to D1.X9 and D16 (3/Fig. 2).
SIREMOBIL Compact
Installation and Setting Instructions
1
Register 3
RXR2-130.033.01
Rev. 04
2
3
2
Fig. 2
Page 4 of 10
04.02
CS PS 24
1
1
1
1