Siemens SIREMOBIL Compact Installation And Setting Instructions Manual page 7

Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

Installation and Setting Instructions
Checking the programmed values
NOTE
Prerequisites
The Qt * Kg value was checked and is programmed. Refer to Programming.
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Calibrations... menu.
Click "Yes" to answer the query.
Select "Diamentor" in the combo box "Function Group".
Select "Start Diamentor Test" in the "Available Functions" window and click "Execute".
Close the status window by clicking on the button "OK".
Answer the query "Do you want to reset the unit" with clicking on the button "No".
Press 2 times the key ->0<- at the control panel of the basic unit (reset key for
fluoroscopy timer).
The tested Qt*Kg value is displayed at the monitor of the SIREMOBIL Compact.
This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: + 3 digits.
By pressing the ->0<- key 2 times, the test is repeated and is displayed again at the
monitor.
The CAL potentiometer on the electronic system of the area dose measurement device
can be used to adjust the test value to the calculated and programmed Qt*Kg value.
Select "Stop Diamentor Test" on the service PC after the adjustment and click "Execute".
Close the status window by clicking on the button "OK".
Answer the query "Do you want to reset the unit" with clicking on the button "Yes".
The test is ended.
Exit the service program with "Logoff" and "Quit".
Check the accuracy of the area dose measurement device. Refer to the next page.
NOTE
Siemens AG
Medical Engineering
Checking the programmed values will activate the self-test in the
electronic system of the area dose measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
Register 3
RXR2-130.033.01
Rev. 04
Page 7 of 10
04.02