Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

Installation and Setting Instructions
Evaluation
Calculate the deviation between the area dose product measured and the area dose
product displayed:
The deviation must be < 0.3.
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ]
/ (measured area dose product FDPg).
For deviations > 0.3, replace the area dose measurement device.
Sirephos
Centering cross
I.I.
Fig. 3
Siemens AG
Medical Engineering
Register 3
RXR2-130.033.01
Rev. 04
Sirephos
Meas. chamber
I.I.
Fig. 4
Fig. 5
Page 9 of 10
04.02
CS PS 24
9
1
SIREMOBIL Compact