Siemens SIREMOBIL Compact Installation And Setting Instructions Manual page 8

Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

8
Checking the accuracy of the area dose measurement device
Prerequisites
The calibration of the area dose measurement device must have already been performed.
Preparations
Secure and center the centering cross on the I.I. (Fig. 1).
Select survey format.
Select kV mA curve plateau HC1.
Release fluoroscopy briefly, adjusting the iris on the radiation field to a format of
approximately 15 cm diameter.
NOTE
Do not make any further adjustments to the iris.
Record the average diameter (Fig. 3) of the octagonal area of the radiation field.
Secure and center the dose measurement chamber of the dose measurement instument
(DIADOS...) on the image intensifier (Fig. 2).
Select kV STOP and enter 70 kV.
Area Dose measurement
Press the >0< key twice to reset the area dose product display on the Monitor to 0.
Reset the dose measurement instrument to 0.
Release fluoroscopy for several seconds. The measured dose K
enough that the uppermost measurement range of the dose measurement instument
(DIADOS...) is utilized fully and the area dose product displayed at the monitor is a
multidigit value (>9); if necessary, reset the area dose product display on the Siremobil
and the dose measurement instrument and repeat the measurement with a different
fluoroscopic time.
Record the measured dose.
Record the area dose product displayed on the montor (FDPa).
Calculating the area dose product
To calculate the measured area dose product:
(measured area dose product FDPg) = (measured dose K
0.785.
SIREMOBIL Compact
Installation and Setting Instructions
The radiation field must cover the entire area of the area dose
measurement chamber.
Register 3
RXR2-130.033.01
Rev. 04
) * (average diameter)