10
Final Steps
•
Switch off the SIREMOBIL.
•
Reinstall all covers.
•
Perform the protective conductor test according to ARTD - 022.731.17.. .The protective
conductor resistance must not exceed 0.2 Ohm.
Changes to previous version
Page 2
Page 6
Page 7
Page 9
SIREMOBIL Compact
Installation and Setting Instructions
Measuring devices actualized.
Corrections in descriptions added.
Corrections in descriptions added.
Maximum values for the protective conductor resistance added.
Register 3
RXR2-130.033.01
Rev. 04
Page 10 of 10
04.02
CS PS 24
1
1
Siemens AG
Medical Engineering