Download Print this page

Final Steps - Siemens SIREMOBIL Compact Installation And Setting Instructions Manual

Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

10

Final Steps

Switch off the SIREMOBIL.
Reinstall all covers.
Perform the protective conductor test according to ARTD - 022.731.17.. .The protective
conductor resistance must not exceed 0.2 Ohm.
Changes to previous version
Page 2
Page 6
Page 7
Page 9
SIREMOBIL Compact
Installation and Setting Instructions
Measuring devices actualized.
Corrections in descriptions added.