Ex pz safety equipment

C.1
Ex pz safety equipment
1
2
3
4
5
Figure C-1
Connect the purge outlet (6) of the purging unit with the purging gas inlet of GA700 (see
Connecting the wall-mounted device (Page 51)).
Connect the purging gas outlet of GA700 (see Connecting the wall-mounted device
(Page 51)) with the purge return (7) of the purging unit.
60L volume of enclosure, multiplied by 10 acc. IEC 60079-2 resulting in 600 L pre-purge
volume. Divide this by the purge flow rate and round it up to the next even number of
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
System status indicator
Air inlet 1/4" BSPP (f)
Flow control
Cable gland for system cable connection 9
System breather – DO NOT OBSTRUCT
Expo Mini purge Type Z (7MB8000 – 1AA)
6
Purge outlet to enclosure 3/8" BSPP (f)
Fitted with 10 mm O.D tube adaptor
7
Purge return from enclosure 3/8" BSPP (f)
Fitted with 10 mm O.D tube adaptor
8
Air outlet – DO NOT OBSTRUCT
Relief valve type RLV25
 
 
C