Technical data
7.1 General technical specifications
Table 7-14
OXYMAT 7: Gas connections
Gas connections
With pipes
Table 7-15
OXYMAT 7: Parts in contact with sample gases
Materials of parts in contact with
sample gases
Sample chamber
Gas path
 
Sealing material
1)
The high-temperature version of the analyzer module is not intended for installation in the rack unit.
Table 7-16
OXYMAT 7: The maximum release amount of sample gas
Analyzer module
OXYMAT 7
7.1.4

ULTRAMAT 7

Table 7-17
ULTRAMAT 7: General technical data
General
Information on the smallest span
Max. power consumption
Weight (only module)
Max. EMC error during operation with one
and/or two OXYMAT 7/ ULTRAMAT 7 units
1)
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
94
Connecting socket for pipe with ∅ 6 mm
OXYMAT 7
Stainless steel:
Plates: Material No. 1.4571
Screw-in glands: No. 1.4571
Hastelloy:
Plates: Material No. 2.4602
Screw-in glands: Material No. 2.4819
Stainless steel:
Pipes: Material No. 1.4404
Gas connections: Material No. 1.4571
Hastelloy:
Pipes: Material No. 2.4602
Gas connections: Material No. 2.4819
FKM (e. g. Viton) or FFKM (Kalrez 6375)
Max. release amount
200 μl/min
ULTRAMAT 7
≤ 30 W
Max. 5.2 kg
OXYMAT 7 high-temperature
Stainless steel:
Plates: Material No. 1.4571
Hastelloy:
Plates: Material No. 2.4602
Stainless steel:
Gas connections: Material No. 1.4571
Hastelloy:
Gas connections: Material No. 2.4602
ULTRAMAT 7 high-temperature
See module nameplate
≤ 120 W
Max. 7.2 kg
1.0% according to EN 61326-1
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
1)
1)