Commissioning

5.1

Basic safety notes

Introduction of toxic and harmful gases into the device
Danger of poisoning
Various gases with compositions known only to the operator can occur in introduced gas
mixtures. The introduction of toxic gas mixtures can result in poisoning.
• Only introduce gas mixtures into the device whose composition is known to you and that
• In the case of toxic, aggressive and/or flammable gases, ensure that the components are
Improper commissioning in hazardous areas
Device failure or risk of explosion in hazardous areas.
• Do not commission the device until it has been mounted completely and connected in
• Before commissioning take the effect on other devices in the system into account.
Danger of poisoning by escaping gas
Leaky gas paths lead to accumulation of the sample gas in the device.
• Check whether the threaded joints at the gas connections are secure. Tighten the threaded
• Check the gas paths for leaks. See section "Checking gas paths for leaks (Page 68)".
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
WARNING
are suitable for analysis in the respective device. Observe the safety information in these
operating instructions. If in doubt, contact your local SIEMENS partner.
suitable for use.
WARNING
accordance with the information in AUTOHOTSPOT.
WARNING
joints with a suitable open-ended wrench, if necessary. In the process, make sure that you
counter properly.
5