6.2.2 RESULTS Mode
Step No.
Setting Method
Remarks
MJ-L7
Display
Step 1
Test mode STOP state
[ tsmmm
A Ace
|
Step 2
Press the MENU button twice
RESULTS mode menu display
[ RESULT ]
Step 3
Press the PLAY/PAUSE button
Display of measurement: Tracking error offset
[ TEO:00 |
Step 4
Press the MENU button
Display of measurement: Focus error offset
[ FEO:00 |
Step 5
Press the MENU button
Display of setting: High reflective focus gain
[HFG: ee]
Step 6
Press the MENU button
Display of setting: High reflective pit tracking gain
[HTG: ee]
Step 7
Press the MENU button
Display of setting: High reflective pit tracking error balance
[HTB: ee]
Step 8
Press the MENU button
Display of setting: Low reflective focus gain
[LFG: ee
Step 9
Press the MENU button
Display of setting: Low reflective pit tracking gain
[LTG: ee
Step 10
Press the MENU button
Display of setting: Low reflective pit tracking error balance
[LTB: ee
Step 11
Press the MENU button
Display of setting: Low reflective total signal level
[GTL: ee]
Step 12
Press the MENU button
Display of setting: Low reflective groove tracking gain
[GTG: ee]
Step 13
Press the MENU button
Display of setting: Low reflective groove tracking error balance
[GTB: ee]
Step 14
Press the STOP button of the Unit
RESULTS menu display state
@
OO: Measurement, @@: Setting, mim: Model code, A:
Microprocessor ROM version (From 01),
& &: EEPROM protect data version (From 01)
When the CANCEL button of DISPLAY unit is pressed in each step, the unit will shift to the opposite step.
When the + button of DISPLAY unit is pressed when the setting is displayed, the setting will increase and the new setting will be stored
in the RAM.
[ RESULT ]
When the — button of DISPLAY unit is pressed when the setting is displayed, the setting will decrease and the new setting will be stored
in the RAM.
When the STOP button of the Unit is pressed when the RESULTS menu is displayed, the test mode STOP state will be set.
When the STOP button of the Unit is pressed when the measurement or setting is displayed, the RESULTS menu display will be set.
6.2.3 MANUAL Mode
Setting Method
Remarks
Test mode STOP state
Display
[ SMM
A ACH & |
Press the MENU button three times
MANUAL mode menu display
[ MANUAL ]
Press the PLAY/PAUSE button
Setting: Temperature
[ TMP:¢¢ ]
Press the MENU button
Setting display: Tracking error offset
[ TEO:00 ]
Press the MENU button
Setting display: Focus error offset
[ FEO:00 |
Press the MENU button
Laser lighting
[LON:
]
Press the MENU button
Shift to the slide innermost circumference
Focus gain rough adjustment (High reflective disc)
Focus gain rough adjustment (Low reflective disc)
[ Hfg:O00@e |
[ Lig:00@e ]
Press the MENU button
Pit: Tracking gain adjustment (High reflective disc)
Pit: Tracking gain adjustment (Low reflective disc)
[HTG:0060e ]
[LTG:000e ]
Press the MENU button
Pit: Tracking error balance adjustment (High reflective disc)
Pit: Tracking error balance adjustment (Low reflective disc)
[HTB:006@e ]
[ LTB:COe@e |
Press the MENU button
Focus gain fine adjustment (High reflective disc)
Focus gain fine adjustment (Low reflective disc)
[HFG:000e ]
LFG:00@e |
Press the MENU button
Press the PLAY/PAUSE button
Moves to step 14 only for the low reflective disc
Moves to step 12
When the PLAY/PAUSE button is
pressed at step 11.
Pit continuous playback
Press the CANCEL button
Moves to step 10
When the MENU button is pressed
at step 11.
TOTAL signal level adjustment
GTL:00e@e |
37