Probe connection
3.3 Probe connection
The probe connections are on the lower right of the
front panel.
Connect straight-beam probes to the red socket (Re-
ceive - right). With T/R probes, connect the pulser cable
to the green socket (left) and the receiver cable to the
red socket (right).
– Set the pulser power as follows:
– Switch the USN 52 R on with K.
– Select the function group RCVR by pressing C.
– Using N or O, set the damping resistor in function
DAMPING There are 4 settings available: 50, 75, 150
and 1000 ohms.
Krautkramer USN 52R/USN 52L
3.4 Starting the USN 52 R
Switching on
This is the normal way of switching on the USN 52 R.
– Press the switch-on key K in the operator's control
panel.
After a short time, the start display appears with infor-
mation about the applied software version and about
the instrument configuration.
The settings for all function values and default settings
are the same as before switching off.
Issue 05, 02/00
Setting into operation
3-5