Epson USN 52R Operating Manual page 88

Lightweight, compact ultrasonic flaw detectors
Table of Contents

Advertisement

Configuring the USN 52R/USN 52L
GRATICULE
Display graticule
In this function you either switch the display graticule
on or off. In doing this, you are able to directly read off
the echo positions and heights.
Operation:
– Highlight the left function group in the left function
table with JC.
– Enable or disable GRATICULE with M. The
currently active setting is displayed.
Krautkramer USN 52R/USN 52L
AMPLITUDE
In this function you determine how the amplitude of the
highest echo is to be displayed in the gate. This ap-
pears in the measurement line as an "H = ..."-value.
You have the following possibilities:
• % Screen HT: A percentage of the screen height
• dB THRESHOLD: (dB REF not active) expressed as
a dB difference relative to the gate threshold
• dB THRESHOLD: (dB REF active) expressed as a
dB difference relative to the stored reference echo
amplitude
Operation:
– Highlight the left function group in the left function
table with I.
– With M select the required setting.
Issue 05, 02/00
Amplitude display
Operation
5-43