Electrolyzed Water Concentration Measuring Tools - Sanyo VW-VF10BG Technical Data & Service Manual

Floor-mounted model with virus washer function
Hide thumbs Also See for VW-VF10BG:
Table of Contents

Advertisement

2. Electrolyzed Water Concentration Measuring Tools

Prepare the following instruments, parts, and materials before starting.
(1) Residual chlorine concentration meter
(Example: Pocket calorimeter produced by Hach)
(2) Total chlorine reagent
(Reagent designated by the manufacturer of the
residual chlorine concentration meter)
(3) Container for measuring the water sample
concentration
Container provided with the residual chlorine
concentration meter
(4) Container for dividing the purified water beaker
(approximately 50 mL)
(5) 10 mL syringe
(for purified water: used to dilute the concentration)
(6) 1 mL syringe
(for the water sample)
(7) Purified water
(Approximately 500 mL of purified water is used to
dilute the water sample. Sterile water for contact
lenses may also be used.)
(8) Bucket
(1 – 3 L: for a drainage container)
3 - 4
Measuring The Water Sample
3