Figure 5-4
The difference between Global View and the well plate image

5-4 Set concentrations or volumes in wells

1.
Click and drag in the well plate image to select the wells where a fluid will be dispensed.
2.
To add the same single concentration or volume of fluid to the selected wells, right-click
anywhere in the well plate image and then click Set Concentration or Set Volume. The Set
Concentration or Set Volume dialog box appears. See
or Set Volume is displayed is determined by the setting chosen in the Edit: Units pane of the
Options dialog box as described in
Select a fluid from the Fluid drop-down list box and then type in a concentration or volume
that is within the valid range that is stated in the dialog box. See
concentration mode.
Figure 5-5
NOTE:
The lowest possible concentration or volume is determined by the single drop
volume, the fluid concentration, and the assay volume. The maximum concentration or
volume is determined by the DMSO limit and the fluid concentration.
―Options: Edit:
Set Concentration dialog box
HP D300 Digital Dispenser
Edit check box
Well plate image
Figure
5-5. Whether Set Concentration
Units‖.
Figure
Valid concentration range
(from one small drop to the DMSO limit)
Global View
5-5, shown for
35