Spare Parts And Accessories; Maintenance And Servicing; Cleaning Sensor Head; Calibration - Beko METPOINT Installation And Operating Manual

Thermal flow meter
Table of Contents

Advertisement

EN
Installation and operating manual

7. Spare parts and accessories

The accessories available for the METPOINT
FLM are listed in the table below.
®
Designation
Picture
Power supply with plug-type connector A

8. Maintenance and servicing

Regularly check the sensor head for dirt and clean it, if necessary. Dirt, dust or oil deposits on the sensor element cause incorrect
measurements.
We recommend checking the sensor element at least once a year. If the compressed air is heavily contaminated, choose a shorter
inspection interval.

9. Cleaning sensor head

To clean the sensor head, immerse it in warm water with a little detergent. Do not clean the sensor with a cloth, sponge, brush or
other implement, as any mechanical impact can destroy the sensor. In the event of persistent deposits, return the sensor to the
manufacturer for inspection and cleaning.

10. Calibration

If the device is not custom-configured, we recommend having it calibrated every 12 months. For calibration, send the METPOINT
®
FLM to BEKO TECHNOLOGIES GmbH.

11. LED indicator

The METPOINT
FLM features a LED calibration indicator mounted at the top of the housing. After 15 months, the LED begins to
®
flash, indicating that the device needs to be calibrated. The flashing LED does not have any effect on the measuring process, and
the device continues to provide accurate measuring signals.
On request, the calibration interval can be adjusted at the factory.
39
METPOINT
SF53
®

Advertisement

Table of Contents
loading

Table of Contents