7. Spare parts and accessories
The accessories available for the METPOINT
Designation
Power supply with plug-type connector
A
4032115
High pressure protection device
4025892
8. Maintenance and servicing
Regularly check the sensor head for dirt and clean it, if necessary. Dirt, dust or oil deposits on the sensor element cause incorrect
measurements.
We recommend checking the sensor element at least once a year. If the compressed air is heavily contaminated, choose a shorter
inspection interval.
9. Cleaning sensor head
To clean the sensor head, immerse it in warm water with a little detergent. Do not clean the sensor with a cloth, sponge, brush or
other implement, as any mechanical impact can destroy the sensor. In the event of persistent deposits, return the sensor to the
manufacturer for inspection and cleaning.
10. Re-calibration
If the device is not custom-configured, we recommend having it calibrated every 12 months. For calibration, send the METPOINT
FLM to BEKO TECHNOLOGIES GmbH.
11. LED display
There is an LED on the top of the METPOINT
15 months to indicate there is an upcoming re-calibration. The flashing LED does not influence measured values. The measurement
signal is still output.
The time interval can be adjusted as requested by the customer in the manufacturing plant.
METPOINT
SF53
®
FLM are listed in the table below.
®
®
FLM housing, which is used to indicate the calibration time. The light flashes after
Installation and operating manual
Picture
®