7. Move the cursor to the Value column and the Temperature Check Table appears, with the cursor
on Entry line 01 (see Figure 4-43). Note the Entry line number when you enter the table so you
can exit the table from the same, if desired.
Entry
[01]
8. On the Temperature Check screen, move the cursor as necessary to select the Zone Name and
to change less than (<), more than (>), alarm value, and the alarm mode of the desired entries.
When you exit the Temperature Check screen, the value in the Entry line
highlighted when you exit becomes the Value entry in the Method Table. Unless
you want to change the Value entry, exit on the same Entry line number that you
entered the Temperature Check screen.
9. Verify the changed values and place the cursor on the desired Entry number. Then press the F1
(Exit) soft key.

Controller Type

There are several types of temperature controllers, such as Sulfur, Sparger, PINA, PNA, and Regular.
This screen is set at the factory and should not have to be changed. If you want to verify the setting,
perform the following steps.
1. On the Background screen, press the F2 (Manual Control) soft key.
2. On the Manual Control Mode screen, press the F4 (Press. & Temp. Control) soft key.
3. On the Pres and Temp Control screen (see Figure 4-39), cursor to TEMP CONFIGURATION and
press the F4 (Controller Type) soft key.
4-47
*****
Temperature Check
Zone
Temp Zone1
02
Temp Zone1
03
Temp Zone1
04
Temp Zone1
05
Temp Zone1
06
Temp Zone1
07
Temp Zone1
08
Temp Zone1
Exit
F1
F2
Figure 4-43. TEMPERATURE CHECK SCREEN
Temperature
C
Action
<
0.0000
Abort
<
0.0000
Display
<
0.0000
Abort
<
0.0000
Display
<
0.0000
Display
<
0.0000
Display
<
0.0000
Display
<
0.0000
Display
More Below
F3
CAUTION
*****
F4
2000-OSM, F1