ABB PGC2000 Operation And Service Manual page 243

Process gas chromatograph
Table of Contents

Advertisement

CALIBRATION STREAM METHOD TABLE
Time
0001
Temp. Contrl
0120
Noise Calc On
0125
Noise Calc Off
0128
Auto Zero N/A
0132
Valve On
0155
Valve Off
0160
Slope Detect
0160
Peak Width Eight
0175
Component One
0175
Component One
0182
Component Two
0188
Component Three
0208
Component Four
0217
Component Five
0217
Baseline
0247
Component Six
0262
Component Seven
0290
Component Eight
0290
Baseline
0312
Component Nine
0312
Baseline
0343
Component Ten
0343
Baseline
0359
Component Eleven
0359
Baseline
0390
Component Twelve
0390
Baseline
0424
Component Thirteen
0424
Baseline
0440
Component Fourteen
0440
Baseline
0462
Component Fifteen
0462
Baseline
0509
Component Sixteen
0509
Baseline
0540
Component Seventeen
0540
Baseline
0570
Component Eighteen
0570
Baseline
0601
Component Nineteen
0601
Baseline
0650
Slope Detect
0710
Stream Step
The various time coded functions are defined as follows. Section 4 contains more complete
descriptions of these functions.
0001
Temp. Contrl
This function ensures that the temperature of the isothermal oven is at the set point value prior to
sample injection.
2000-OSM, F1
Method Table #10
Function
Value
One
One
One
On Now
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Force @ Valley
Off
N/A
One
7-29