Siemens SIWAREX FTA Device Manual page 141

Weighing electronics for automatic scale
Table of Contents

Advertisement

Code
Command explanation
Characteristic curve shift
15
The instruction shifts the entire characteristic curve
(adjustment coordinates) so that the current filtered
digital value from the ADC is made into a zero
point. Only permitted with a calibration switch that
is not set. If an impermissible digital value results
from the shift then the instruction is rejected.
Scale commands
Zeroing
21
The current weight is set to zero. In calibratable
operation ("OIML"), only possible with restrictions
(-1%, +3%). The tare is deleted at the same time.
22
Tare
The current weight is set to zero and the weight
display is designated as "Net" and "Tare" at the
same time.
Delete tare
23
The tare is deleted. The current weight is displayed
and designations of "Net" are changed to "Gross",
the designation "Tare" or "Preset-Tare" is reset.
24
Accept tare entry
The defined tare is accepted as tare and at the
same time, "Preset tare" is designated together
with the weight display.
Switch on increased resolution
25
Activate output / display of the calibratable weight
value with increased resolution for 5 seconds.
26
Display tare weight
Activate output / display of the tare value for 5
seconds.
29
Display firmware
Displays the firmware version in the DR34 for 5
seconds
SIWAREX FTA
Commands
Executable
in following
operating
states
Service
operation
No weighing
cycle, Service
operation
No weighing
cycle
No weighing
cycle
No weighing
cycle
All
All
All