Error
Data and
No.
command errors -
description
89
90
91
92
93
94-
Scale command
104
cannot be
executed
94
Adjustment weight
too small
95
Reserved
96
Tare T exceeded
97
Totalising memory
1 deletion not
permitted
98
Totalising memory
already deleted
99
Totalising memory
1 could not be
logged
100
Weight outside zero
definition range
101
Scale command
cannot be executed
since not at
standstill 1
102
Adjustment weight
missing
103
Characteristic curve
shift not possible
104
SIWAREX FTA
Solution
Reserved
Reserved
Reserved
Reserved
Reserved
The adjustment weight in DR 3 is too low. Increase the difference
to the adjustment weight. The measurement value between
sequential adjustment weights is to be at least 5 % of the
measurement range of the load cells apart
Reserved
The subtractive Tare has been exceeded. This error is generated in
the following cases:
With the Tare command, the gross value lies over the defined
permitted tare range.
An externally defined tare value is negative.
An externally defined tare value is greater than the max-value of
weighing range 1 on a multi-segment scale.
An externally defined tare value is greater than the defined max tare
value [%] of the largest weighing range in calibratable operation
(specifications Code OIML in DR 3) of a multi-range scale.
The distribution memory cannot be deleted.
The command "Delete distribution memory 1" (114) is not
permitted in calibratable operation. Use command (113) -
"Delete/Log calibratable distribution memory 1".
The command "Delete calibratable distribution memory 1" (113)
can only be terminated in calibratable operation (Specifications
Code OIML in DR 3) if log text 1 (DR40) contains the distribution
memory 1 field.
Check the parameters.
The delete totalizing memory command could not be executed
since the memory is already deleted.
Logging the distribution memory has failed. Check the output
device. Check additional technological faults for indications as to
the cause of the fault. A technology fault is generated indicating
the cause.
Zero setting cannot be performed since the current zero point lies
outside the defined zero setting range in DR 3 or the current
gross value lies above the largest defined weighing range value.
The weighing command (Taring, Zero setting, Logging) cannot
be executed because Standstill 1 does not lie within the standstill
time defined in DR 3.
Adjustment command cannot be executed because the
respective adjustment weight in DR 3 has not been entered.
Adjustment instruction for characteristic curve shift cannot be
executed. Possible causes: Shift leads to exceeded range,
adjustment not completed, write protect is active).
Reserved
Messages and Diagnostics
7-145