Equipment Required - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

Equipment Required

R11
Gain
2264 GAX/HAX/JAX/KAX/LAX/MAX
7-24
Calibration: Nautilus Tocotransducer Calibration
You will need the following:
Phillips Head Torque Driver
Screw Cap Extraction Tool
Digital Voltmeter: DM501 or equivalent with x1 probe or leads
Common Mode Rejection (CMR) Test Jack
Potentiometer Adjustment Tool
52.5 gram weight
NOTE: You will need a specific weight to calibrate the Nautilus
Tocotransducer. Refer to
part number information.
Torque Seal
Loctite Adhesive #454 (recommended)
Anti-Static Wristband
Static-Free Work Surface
Calibrated 120 Series Monitor
Flat Screw Cap
Button-Style
Figure 7-6. Sealing Screw Caps
R6
Offset
Gain
Figure 7-7. Nautilus Tocotransducers—Cover Removed
120 Series Maternal/Fetal Monitor
2015590-001
(Figure
"Chapter 14, Nautilus Tocotransducer"
Loop-Style
R15
2264 AAX/BAX/CAX/DAX/EAX/FAX
7-4)
for
R5
Offset
R6
Offset Trim
Revision B