Transmitted Data Format

Monitor Type

Response Type

Monitor ID

Data Field

Annotations

Revision B
Peripheral Devices: Model 115-Compatible Communications Protocols
The format of the information transmitted from the 120 Series Monitor to the
external computer is shown in
NOTE: This applies to both 115 Update mode and 115 Transmit/Receive mode.
The definitions of the bytes are as follows:
This field contains a 1-byte ASCII value indicating the type of monitor sending the
information. The 120 Series Monitor is set to a fixed value of 9 (39H).
This 1-byte value indicates the type of information being sent from the 120 Series
Monitor as shown in
This 1-byte ASCII value indicates the ID number of the monitor information. The
120 Series Monitor sets this byte to a fixed value of 0 (30H).
This field of n-bytes contains the actual data sent from the monitor to the external
computer. The contents of this field for each response is given below:

Event Mark

This data field is empty whenever the monitor's

Heart Rate

This data field contains two or three ASCII characters representing the value of the
heart rate. A value of 00 (30H, 30H) indicates a penlift condition. Heart rate data
(including penlift data) is transmitted four times per second.

Uterine Activity

This data field contains two or three ASCII characters representing the value of
uterine activity. A value of 128 (31H, 32H, 38H) indicates a penlift condition.
Uterine activity data (including penlift data) is transmitted eight times per second.

Modes

This data field contains two ASCII characters indicating the mode. The first
character is the combined mode of both heart rates according to
second character is the mode for uterine activity as listed in
transmitted at startup, with each mode change, with each request, and once every
minute.
This field of n-bytes contains actual ASCII data making up messages sent to the 120
Series Monitor. This information can be: messages sent from a Model 2115/2116B
Data Entry System; or patient parameter information (SpO
monitor's internal modules or external devices connected to the monitor.
120 Series Maternal/Fetal Monitor
2015590-001
Figure
11-3.
Table
11-1.
Mark
pushbutton is pressed.
Table
11-2. The
Table
11-3. Mode data is
and/or NBP) from the
2
11-11