Patient-to-Line Leakage for FSpO
Revision B
General Maintenance: Maternal Blood Pressure Tests
2
1.
Connect an FSpO
test body to the monitor's front panel
2
assembly leakage cable for the test body.
Leakage
2.
Set the fixture's
3.
Turn ON the monitor using the monitor's power switch.
4.
Verify the following:
< 50 µA (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-10. Patient-to-Line Leakage Conditions for FspO
Neutral
Closed
Closed
Open
Open
120 Series Maternal/Fetal Monitor
2015590-001
Line
switch to
.
Ground
Power (Polarity)
Closed
Closed
Closed
Closed
FSpO
input. Use the
2
2
Normal
Reversed
Normal
Reversed
9-21