Response Time; Print Interval; Fspo2 Setup Screen - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

FSpO
Setup Screen
2

Response Time

Print Interval

Revision B
Setup Procedures: FSpO2 Setup Screen
FSPO
Select the
softkey to access the FSpO
2
IMPORTANT
SOFTKEY ACTIVATION—The Corometrics fetal patient
module cable must be plugged into the
FSpO
activate the
mode softkey.
2
FECG
165 172
FSpO
RESPONSE TIME: SLOW
PRINT INTERVAL: 5 MIN
% O
TRACE: OFF
2
Figure 5-9. FSpO
Setup Screen
2
This field sets the response time or averaging mode (the time the monitor takes to
respond to changes in fetal oxygen saturation).
Slow: The FSpO
display responds to step changes in the fetal saturation with
2
approximately 120 accepted pulses, or in about 50 seconds at 150 BPM.
Fast: The FSpO
display responds to step changes in the fetal saturation with
2
approximately 30 accepted pulses, or in about 11 seconds at 150 BPM. (The
factory default setting is SLOW.)
When signal quality is poor, the response time becomes longer regardless of the
mode selected.
This setting determines the time interval for printing the FSpO
chart paper.
120 Series Maternal/Fetal Monitor
2015590-001
Setup screen
2
FSpO
connector in order to
2
US
TOCO
30
SETUP
2
FSpO
2
47%
EXIT
(Figure
6-2).
values on the strip
2
5-19

Advertisement

Table of Contents
loading