Maternal Blood Pressure Tests; Accessing The Diagnostic Control Screen - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

Maternal Blood Pressure Tests

Accessing the Diagnostic Control Screen

NOTE: The correct date and time must
be set on the General Setup screen or
you will not gain access to the service
screens.
9-6
General Maintenance: Maternal Blood Pressure Tests
The manometer and pneumatic pressure tests are initiated from one of several
service screens:
SETUP
1.
Select the
softkey from the main operating screen to display the General
Setup screen
(Figure
SERVICE
2.
Select the
softkey from the General Setup screen to display the Service
Lock screen
(Figure
3.
Use the Trim Knob control to set the access code to the current month and date.
You can enter the date followed by the month; or the month followed by the
date. For example: February 21 can be entered as
0 2 2 1
2 1 0 2
or
.
4.
As soon as you enter the correct access code, the Install Options screen displays
(Figure
9-3).
120 Series Maternal/Fetal Monitor
2015590-001
9-1).
9-2). The access code is displayed as
0 0 0 0
.
Revision B