Connecting
4.1 Electrical connections
4.1.1.2

Wall-mounted device

Cable glands
The cables must have the correct diameter for the cable glands. After installation, make
sure that the cables are positioned securely and that the seals are in the correct position.
Danger due to electrical current
Check that the existing power supply corresponds to that specified on the nameplate of the
device.
Danger due to electrical current
The power supply cable to be connected at the customer location could be damaged by
incorrect fitting in the device.
You should therefore carry out a visual inspection and PE conductor test of the cable in
accordance with the country-specific safety standards, regulations or other applicable
directives before starting up the device. Since this is a cable which lies partially within the
device, you must always additionally apply the production safety standard IEC / EN 61010-
1 or its national version.
4.1.2

Rack-mounted device

4.1.2.1

Electrical connection of the power supply

Procedure
Note
Appliance plug
You can use any appliance plug that fulfills the conditions of DIN EN 60320-1, ≥6 A.
The mains supply line must comply with the valid regulations and conditions for the place of
installation and be provided with a protective conductor at the potential of the housing. The
cross-section of each wire has to be ≥ 1 mm
in the connector at the position marked.
18
WARNING
WARNING
WARNING
. The phased terminal lead must be connected
2
Compact Operating Instructions, 01/2013, A5E31805656-01
Quick Start