Settings > Gas Path/Process Tag Label; [2.12] Settings > Gas Path/Process Tag Label - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

7.9
[2.12] Settings > Gas path/process tag label
Note
Parameter assignment of the gas path/process tags
The parameter assignment of the process tags in case of a separate gas path is only
possible if two analyzer modules are installed in the device. If there is only one analyzer
module, this setting option is not available.
Requirement
Before you configure the gas path of the device, all lines conducting gas must be correctly
connected.
For more information about connecting the device, go to the list of references in Table A-3
LVZ3 (Page 81).
Procedure
1. Component > Main menu > "2 Settings" > "12 Gas path/process tag label"
2. Set the design: "Gas path" parameter field.
3. Enter the process tag labels:
See also
References (Page 81)
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
Authorize access with Expert PIN, if necessary.
Specify the design:
– "Shared" (standard for devices with only one analyzer module)
– "Separate" (one gas path for each of the installed analyzer modules)
If you have assigned a shared gas path:
– Open the "Designation of process tag" parameter field.
– Enter a designation of up to seven characters in the text box.
If you have assigned a separate gas path:
– Open the "Process tag ..." parameter field.
– Enter a designation of the module-specific process tag of up to seven characters in
the text box.
– Repeat the process for the process tag of the second analyzer module.
The process tag label is displayed in the title bar.
7.9 [2.12] Settings > Gas path/process tag label
Commissioning
69