Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
3
Press Channel 1 50Ω/1MΩ key.
The LED turns on, indicating a 50Ω input impedance for Channel 1.
4
Press Channel 1 DC/AC key.
The LED turns on, indicating DC input coupling for Channel 1.
5
Verify the DVM reads 51.1Ω ± 3Ω.
6
For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 5a.
6'
For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 4a.
7
Press Channel 1 50Ω/1MΩ key.
The LED turns off, indicating a 1 MΩ input impedance for Channel 1.
8
Verify the DVM reads approximately 1.0000000 MOHM.
9
For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 5b.
9'
For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 4b.
10
Repeat steps 3 through 9 for Channel 2 for the HP 53131A/132A
Counter ONLY since the HP 53181A Counter does not have a
standard Channel 2.
11
For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Tests 6a and 6b.
12
Disconnect the test setup.
Assembly-Level Service Guide
1-9