Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
Procedure
1
Set the Pulse Generator to the following:
PERIOD: 20 µs
WIDTH: 5 µs
High: 3.00 V
Low : 0.8 V
Trailing/Leading : 1.0 ns
Input Mode: TRIG
2
Connect the equipment as shown in Figure 1-3.
50Ω
Feedthrough
Ext Arm
(rear panel)
Counter
Figure 1-3. External Arm Test Setup (HP 53131A/132A Only)
Assembly-Level Service Guide
HP 8130A
Pulse Generator
HP 8663A
Synthesized Signal Generator
Channel 1
Output
Output
N-to-BNC
Connector
1-11