Procedure - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)

Procedure

1
Connect the pulse generator signal to Channel 1 of the Counter as
shown in Figure 1-5.
HP53131A/132
Counter
Channel 1
Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup
2
Set the Pulse Generator to the following:
PERIOD: 10 µs
WIDTH: 150 ns
High: 1.00 V
Low : 0.00 V
Input Mode: Normal
3
On the Counter, Press Single/Stop key.
The HP 53131A Counter should display 0.1500 µs ±0.0040 µs.
The HP 53132A Counter should display 0.1500 µs ±0.0025 µs.
4
Record this value in the appropriate place in the Performance
Test Record (Test 1, Line 1 for HP 53131A or Test 1, Line 2 for
HP 53132A).
5
Disconnect the test setup.
1-18
HP 8130A
Pulse Generator
Assembly-Level Service Guide
Output