Test 2: Channel 1 Frequency Sensitivity (Hp 53181A Only) - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 1 Performance Tests

Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)

Test 2: Channel 1 Frequency Sensitivity
(HP 53181A Only)
This set of tests verifies frequency sensitivity specifications of the
HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000 V
50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Assembly-Level Service Guide
1-51